OptiFiber Pro custom test limit: Difference between revisions
No edit summary |
No edit summary |
||
Line 109: | Line 109: | ||
| Splice Loss (dB) | | Splice Loss (dB) | ||
| 0,30 | | 0,30 | ||
| 0,15 or 0.30* | | 0,15 (UPC) or 0.30 (APC)* | ||
|- | |- | ||
| Reflectance (dB) | | Reflectance (dB) | ||
Line 128: | Line 128: | ||
|} | |} | ||
(*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. | (*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. In case of APC terminated fiber, we recommend using the same limit for connection and splice i.e. 0.30dB. | ||
| | ||
Line 146: | Line 146: | ||
| Splice Loss (dB) | | Splice Loss (dB) | ||
| 0,30 | | 0,30 | ||
| 0,15 or 0.30* | | 0,15 (UPC) or 0.30 (APC)* | ||
|- | |- | ||
| Reflectance (dB) | | Reflectance (dB) | ||
Line 165: | Line 165: | ||
|} | |} | ||
(*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. | (*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. In case of APC terminated fiber, we recommend using the same limit for connection and splice i.e. 0.30dB. | ||
| |
Revision as of 09:25, 14 June 2018
Introduction
Length/loss limits to be used with ISO/IEC14763-3:2014 (Information technology -- Implementation and operation of customer premises cabling -- Part 3: Testing of optical fibre cabling) are no longer adequate to evaluate the quality of a recent installed fiber optic infrastructure as they are too relaxed.
Solution
We recommend this custom limit during OTDR testing. End-users could set a Maximum Length and Maximum Overall Loss in accordance with the application limit. For example, if the link is OM3 multimode fiber and the application is IEEE 10GBASE-SR, these values would be respectively 300m and 2,60dB (@ 850 nm). More information is available during the Certified Cabling Test Training module VERSIVOTDR.
Maximum length
Parameter | Default value | Recommended value |
Length (m) | N/A | N/A |
Limits for 850 nm
Parameter | Default value | Recommended value |
Connector Loss (dB) | 0,75 | 0,30 |
Splice Loss (dB) | 0,30 | 0,15 |
Reflectance (dB) | -35,0 | -35,0 |
Overall Loss (dB) | N/A | N/A |
Overall ORL (dB) | N/A | N/A |
Segment Attenuation Coefficient (dB/km) | N/A | 3,00 |
Limits for 1300 nm
Parameter | Default value | Recommended value |
Connector Loss (dB) | 0,75 | 0,30 |
Splice Loss (dB) | 0,30 | 0,15 |
Reflectance (dB) | -35,0 | -35,0 |
Overall Loss (dB) | N/A | N/A |
Overall ORL (dB) | N/A | N/A |
Segment Attenuation Coefficient (dB/km) | N/A | 1,50 |
Limits for 1310 nm
Parameter | Default value | Recommended value |
Connector Loss (dB) | 0,75 | 0,30 |
Splice Loss (dB) | 0,30 | 0,15 (UPC) or 0.30 (APC)* |
Reflectance (dB) | -45,0 | -45,0 |
Overall Loss (dB) | N/A | N/A |
Overall ORL (dB) | N/A | N/A |
Segment Attenuation Coefficient (dB/km) | N/A | 1,00 |
(*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. In case of APC terminated fiber, we recommend using the same limit for connection and splice i.e. 0.30dB.
Limits for 1550 nm
Parameter | Default value | Recommended value |
Connector Loss (dB) | 0,75 | 0,30 |
Splice Loss (dB) | 0,30 | 0,15 (UPC) or 0.30 (APC)* |
Reflectance (dB) | -45,0 | -45,0 |
Overall Loss (dB) | N/A | N/A |
Overall ORL (dB) | N/A | N/A |
Segment Attenuation Coefficient (dB/km) | N/A | 1,00 |
(*) When testing APC terminated fibers, the OTDR will interpret the APC connection as a splice and apply the splice limit accordingly. In case of APC terminated fiber, we recommend using the same limit for connection and splice i.e. 0.30dB.