Battery status: Difference between revisions

From wiki.comcert.com
Jump to navigation Jump to search
No edit summary
No edit summary
 
Line 2: Line 2:
== Introduction ==
== Introduction ==


Multiple uses have noticed unreliable test results when the batteries of the Versiv are below 50%.
Multiple users reported unreliable OLTS test results when the battery of a Versiv unit is below 50% of maximum load.  The stability of the light source could be impacted by a low battery.


 
 
Line 10: Line 10:
The phenomenon has been confirmed by Fluke Networks TAC and engineering is working on a fix. 
The phenomenon has been confirmed by Fluke Networks TAC and engineering is working on a fix. 


Meanwhile, keep the batteries of the Versiv charged at 50% or higher before starting any test.
Meanwhile, keep the batteries of the Versiv charged at 50% or higher before setting the reference or starting a Loss measurement.


 
 

Latest revision as of 11:30, 31 January 2021

Introduction

Multiple users reported unreliable OLTS test results when the battery of a Versiv unit is below 50% of maximum load.  The stability of the light source could be impacted by a low battery.

 

Solution

The phenomenon has been confirmed by Fluke Networks TAC and engineering is working on a fix. 

Meanwhile, keep the batteries of the Versiv charged at 50% or higher before setting the reference or starting a Loss measurement.